Show simple item record

dc.contributor.authorLechuga-Noa, Víctor E.
dc.contributor.authorRodríguez-Tanta, L. Yesenia
dc.contributor.authorGálvez Caballero, David G.
dc.date.accessioned2023-10-26T14:09:27Z
dc.date.available2023-10-26T14:09:27Z
dc.date.issued2023-06-03
dc.identifier.citationRevista del Cuerpo Médico Hospital Nacional Almanzor Aguinaga Asenjo. 2023:16(1).es_PE
dc.identifier.issn2227-4731
dc.identifier.urihttps://hdl.handle.net/20.500.12959/4332
dc.description.abstractIntroducción: El ductus arterioso persistente (DAP) es una cardiopatía congénita de especial interés. La oclusión intervencionista es el tratamiento de elección, no obstante, el dispositivo médico “oclusor para DAP”, usado en este procedimiento, puede dar lugar a la aparición de eventos adversos de rara frecuencia como hemorragia, trombocitopenia, entre otros. Reporte de caso: Se presenta un caso de una paciente pediátrica con trombocitopenia severa posterior a la intervención. Durante su hospitalización, la paciente mantuvo niveles bajos de plaquetas hasta que logra su recuperación y sale de alta sin una nueva manifestación hematológica. El caso fue reportado como sospecha de incidente adverso asociada al oclusor para DAP. Conclusión: Al realizarse el análisis causa-raíz, se concluye que el tamaño del oclusor para DAP podría ser la principal causa de la trombocitopenia. El presente caso es una muestra que los dispositivos médicos pueden causar eventos adversos severos, por lo que, es necesaria la implementación de tecnovigilancia, principalmente a los dispositivos médicos de alto riesgo.es_PE
dc.description.abstractIntroduction: Patent ductus arteriosus (PDA) is a congenital heart condition of special interest. Interventional occlusion is the treatment of choice; however, a PDA occlude, used in this procedure, is a medical device that rarely could lead to adverse events such as hemorrhage, thrombocytopenia, etc. Case report:We present a case of a pediatric patient with severe thrombocytopenia after surgery. During her hospitalization, platelet levels remained lower and, finally, she was fully recovered without any hematological manifestation. The case was reported as a suspected adverse incident related to PDA occlude. Conclusion: After performing a cause-root analysis, we concluded that the size of the PDA occlude may be the main cause of thrombocytopenia. This case demonstrates that medical devices could lead to adverse events. Thus, it is important for clinicians to implement medical device surveillance, mainly for high-risk medical deviceses_PE
dc.formatapplication/pdfes_PE
dc.language.isospaes_PE
dc.publisherSeguro Social de Salud (EsSalud)es_PE
dc.relation.urihttp://cmhnaaa.org.pe/ojs/index.php/rcmhnaaa/article/view/1719es_PE
dc.rightsinfo:eu-repo/semantics/openAccesses_PE
dc.rights.urihttps://creativecommons.org/licenses/by-nc-sa/4.0/es_PE
dc.subjectOclusor DAPes_PE
dc.subjectTrombocitopeniaes_PE
dc.subjectEvento adversoes_PE
dc.subjectTecnovigilanciaes_PE
dc.subjectPatent ductus arteriosuses_PE
dc.subjectPDA occluderes_PE
dc.subjectThrombocytopeniaes_PE
dc.subjectMedical devices Surveillancees_PE
dc.titleTrombocitopenia severa inducida por un oclusor para DAP: reporte de caso en tecnovigilanciaes_PE
dc.title.alternativeSevere thrombocytopenia induced by a occlude for PDA: A medical device surveillance case reportes_PE
dc.typeinfo:eu-repo/semantics/articlees_PE
dc.subject.ocdehttps://purl.org/pe-repo/ocde/ford#3.02.04es_PE
dc.identifier.doihttps://doi.org/10.35434/rcmhnaaa.2023.161.1719


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record

info:eu-repo/semantics/openAccess
Except where otherwise noted, this item's license is described as info:eu-repo/semantics/openAccess

El repositorio recopila la documentación y publicaciones institucionales, producto de la investigación y el desempeño en defensa de la competencia, la propiedad intelectual y protección al consumidor, para su difusión en el entorno social y académico. Se difunde en acceso abierto y está alineado con las Declaraciones de Budapest y de Berlín.

Contactos